MICLAB-110 Laboratory Investigation and Retest Procedure for Atypical and Out of Specification Results

DepartmentQuality ManagementDocument noMICLAB 110
TitleMicrobiology Laboratory Investigation and Retest Procedure for Atypical and OOS Results
Prepared by: Date: Supersedes: 
Checked by: Date: Date Issued: 
Approved by: Date: Review Date: 

 

Document Owner

Micro Laboratory Manager

Affected Parties

All Microbiology Laboratory colleagues

Purpose

The purpose of this procedure is to provide guidance when investigating microbiology laboratory out of specification (OOS) results associated with raw material samples, in-process samples and finished product samples. This procedure describes the actions taken by Microbiology Laboratory staff in the event the result of a test does not conform to company specifications for microbiological release.

This procedure will also provide guidance for re-testing raw material samples, in-process samples and finished product samples when it has been decided through a laboratory investigation that retesting is justified. Retesting should be viewed as an investigational tool to aid in determination of the root cause of the discrepant laboratory result.

Scope

This procedure applies to all microbiological testing used for supporting release of product. This includes non-sterile and sterile in process testing and finished product testing including biological assays like endotoxin testing. This procedure also applies to water testing and disinfectant testing OOS results. Where confirmation of a valid OOS result leads to a more detailed investigation of the manufacturing process or environment, the laboratory investigation form (Form 690) should be attached to the investigation report.

Exclusions from this procedure include, air and surface Environmental monitoring (both viable and non-viable OOS results), Personnel monitoring excursions and process validation failures (eg media fills and soup tests). These excursions should be addressed individually through a Deviation Reporting System (DR) (see QMS 035) or a detailed investigation report.

It is the responsibility of the Laboratory Manager and the delegates to ensure that this procedure is followed in a timely manner, without preconceived assumptions as to the cause of the Out-of-Specification (OOS) result.

It is the responsibility of Microbiology staffs to notify the Microbiology Manager of a deviation or OOS results and ensure this procedure is followed in a timely manner.

An OOS result does not necessarily mean the batch under investigation fails and shall be rejected.  The OOS results shall be investigated and the findings of the investigation, including re-test results if justified, shall be interpreted to evaluate the batch and reach a decision regarding release or rejection.

Definition

Out of Specification (OOS)Out-of-Specification test result is one that lies outside the specifications to which it is required to conform.
Out of Limit (OOL)Out-of-Limit result is a result that exceeds a pre-determined alert limit, which is applied to a sample type in order to assess whether a process is in control. The result exceeds alert limits.
Atypical Test result (ATR)A test result that based on trending data, falls outside of the historical range for the sample/product but is within sample/ product specifications. The result passes but may indicate process drift.
Deviation Report (DR)Documentation system for recording, investigation and analyzing material and processes that do not comply with company requirements.
Known Laboratory Error (KLE)This is a laboratory deviation that is known to be caused by technician or instrument failure. A KLE, which occurs during test set-up and prior to generation of the results, does not require a DR or Form 690.

 

Investigational TestingTesting conducted to assist in the determination of the root cause of and OOS or ATR. Investigational testing must be given prior approval by Microbiology Manager and documented on Form 690 and in DR is appropriate. Results from investigational testing can only used to determine whether the OOS result is valid.
Repeat TestThis is a test that is designed to replace a valid test that has been shown to be invalid, through the investigation process described in this procedure. Once an assignable cause has been established, the discrepant result is invalidated and a repeat test is performed using the original test procedure and the original sample whenever possible.
Investigational TestingTesting conducted to assist in the determination of the root cause of an out of specification or atypical result. Investigational testing must be documented and approved by Microbiology Manager. Results from investigational testing are only used to determine whether the OOS result is valid.
RetestAdditional testing performed using an approved retest protocol as outlined below.
BPNBatch Production Number
BSBBulk Solution Bioburden Samples
FCBFilled Container Bioburden Samples
RMRaw Material

 

Related Documents

Form 690Microbiology Out of Specification (OOS) Investigation and report Form
MICLAB 060Micro Laboratory Procedure for Sterility Testing
MICLAB 085Bacterial Endo Toxin Testing kCA Method
MICLAB 075Micro Evaluation on Bioburden, Non sterile and Raw Materials
MICLAB 080Bacterial Endo Toxin Testing (LAL) – Gel Clot Method
MICLAB 055Microbiological Monitoring of Plant Water Systems
QMS 035Deviation Reporting System

 

EHS Statement

Not applicable

Procedure

1. Timelines

1.1         The Laboratory OOS investigation Form 690 must be initiated within 1 working day of generation of the original OOS result.

1.2         Section A and Section B of Form 690 should be completed within 1 working day of generation of the OOS result. The purpose Phase 1 of Form 690 is to confirm or invalidate the original OOS result as soon as possible and to allow for immediate re-testing or confirmatory testing if deemed appropriate.

1.3         A DR should also be raised within 1 working day if required after the Phase 1 Investigation has been completed. A DR is required if, after completion of Section A and Section B of Form 690 the OOS result is valid.

1.4         Form 690 should be completed with 30 working days. This will allow for further re-testing or confirmatory testing to take place, including microbial identification work if required.

2. Determination of an OOS/OOL Result

2.1         For a result to be Out-of-Specification (OOS), or Out of Limit (OOL) a specification or alert limit must be applied. If a result obtained is determined to be outside the pre-defined limit for the sample being tested, the testing Technician in conjunction with an appropriate delegate must initiate a Phase 1 Laboratory Investigation.

2.2         The Microbiology Technician is responsible for:

  • Being aware of potential problems that may arise during testing and monitoring potential problems that may result in inaccurate or invalid results.
  • Ensuring all test acceptance requirements are met.
  • Reporting an OOS/OOL result to the Microbiology Manager.
  • Investigating the out of specification result by initiating Form 690 and a Phase 1 Laboratory investigation.

2.3         Upon completion of Section A and Section B of the form, approval must be sought from Laboratory Manager.

2.4         If the OOS result is found to be valid, the Microbiology Technician should raise a Deviation Report (DR) for the OOS Investigation immediately and document the DR number on Section A of Form 690.

2.5         At this time, an initial evaluation for further testing is undertaken (Section C of OOS Report and Investigation Form) and deemed to be appropriate or otherwise.

2.6         If a determinant error is found, it is explained in Section B of Form 690 and the result will be invalidated. In this instance a re-test of the original or retention sample can be conducted is the quality of the re-test samples is deemed to be representative of the original sample and the time of sampling.

2.7         If no determinant error is found, confirmatory testing may be performed in order to confirm the out of specification result if deemed appropriate. Prior approval for confirmatory testing by Microbiology Manager is required.

NOTE: When considering whether re-testing or confirmatory testing is valid, any factors that may potentially affect the sample quality should be considered, including but not limited to the age and storage conditions of the original or retention sample and whether the sample type has inherently antimicrobial characteristics. In some cases, re-testing a retention sample may be of no value, and other criteria should be assessed in the context of product release.

2.8         Repeat testing or confirmatory/ investigation testing is initiated only when approval from Micro Manager has been granted.

2.9         Where repeat testing, re-testing or confirmatory testing is conducted, Section C of  Form 690 must be completed, including the number of replicates used for testing, the acceptance criteria for the additional test results and the justification and reason for conducting the additional testing. When re-testing refer to Section 5 for determination of the re-test protocol.

2.10       Results from other testing are recorded in Section C of Form 690.

2.11       If a batch or partial batch is rejected as a result of the out of specification investigation, a level 1 Investigation should be carried out.

2.12       Included in Phase 2, Section C of Form 690 is the batch disposition as well as the corrective and preventative actions when applicable.

2.13       The action plan is reviewed by the Laboratory Manager, or the delegates and the Technician.

3. Initiating Form 690

3.1         For each Action or Alert Limit excursions within the scope of this procedure, a Microbiology Laboratory OOS/OOL Investigation and Report Form 690 is to be initiated.

3.2         Each OOS/OOL Investigation Report form must be assigned a unique identifying number. This number should be referenced in any relevant DR or Investigation Reports associated with the OOS/OOL result.

3.3         The OOS/OOL Investigation report form should be attached to the DR reference where appropriate, and filed in the Microbiology Laboratory Investigations file.

4. Laboratory Investigation Process

4.1          General

4.1.1      In the initial instance of recording an OOS/OOL result, notify Microbiology Manager and/or appropriate delegates.

4.1.2      A test may be repeated when the Initial test has been invalidated due to an error in procedure, (including Technician or equipment failure, or test acceptance criteria not met).

4.1.3      Microbiology Laboratory Investigations will be conducted in two phases – Phase 1 and Phase 2 as detailed below. Phase 1 is the review of the testing procedures by the microbiology technician to eliminate the possibility of laboratory introduced error or contamination. Phase 2 will investigate the OOS result in more detail.

4.1.4      Under normal conditions Alert Level excursions (OOL) will only require Phase 1 and if result is deemed valid, microbial identification and simple trend reports should be conducted. No DR is required for OOL results. Specific Test procedures should provide guidance on when OOL results should be more thoroughly investigated including when to raise DR.

4.1.5      For Action Level excursions (OOS) it is necessary to investigate thoroughly to determine a definitive cause and recommend corrective and preventative actions, (CAPA). If an OOS result, is deemed valid, A DR should be raised in accordance with QMS 035 and further investigated should be conducted.

4.1.6      Phase 1 and 2 should include assessment of the following details as a guideline. This list is not exhaustive and other relevant information should also be included where possible.

4.2          Phase 1

4.2.1       When completing Phase 1 of Form 690 for OOS/OOL results the following details should be evaluated.

  • Sample collection, storage and preparation
  • Sample and test preparation
  • Test equipment (calibration)
  • Technician Training records
  • Expiry of consumables
  • Media quality control passed
  • Test method followed
  • Observations made during test procedure.
  • All calculations checked and verified
  • All test control results, where relevant, are satisfactory
  • Review of other test results for the test session, to determine a trend

 

4.2.2       At the conclusion of Phase 1 Investigation, an evaluation of the validity of the original OOS/OOL result should be made. If through the investigation process, it has been demonstrated that the OOS/OOL result is valid, a DR should be raised and a more detailed investigation initiated, including a root cause analysis.

4.3          Phase 2

4.3.1      Phase 2 involves a thorough systematic analysis of the incident, to determine the root cause for the OOS result. The investigation should be tailored for each incident to reflect the significance of meeting or exceeding either the ALERT or ACTION Levels.

4.3.2      The minimum requirements for most situations are listed as below, this list is by no means exhaustive and specialist assistance from departments other than Microbiology should be utilised in the investigation.

4.3.3      Where re-testing, repeat testing or confirmatory or investigational testing is required to confirm the root cause, refer to Section 5.

5. Retest Procedures

5.1          Prior to conducting any re-testing, Microbiology Manager or delegate must be consulted.

5.2          As part of Phase 2, further testing may be required to determine a root cause(s). Further testing may also be required if a possible cause has been identified.

5.3          A second technician may perform retesting where appropriate or required.

5.4          Standard Retest Protocol

5.4.1      The retest protocol generated in Phase 2 must specify which retest results will be reported

5.4.2      Non-Sterile Bioburden Samples (RM, FCB, BSB, Water) – Perform a retest in triplicate on the original sample and on a retain sample if available.

  • Where original result has been invalidated (Phase 1), re-test of original sample, using original sample size should be conducted.
  • If retests of both original and retain sample confirm the original result, report the initial result.
  • If retest do not confirm the initial result, document results of the retests on Form 690 (Phase 2, Section C); no further testing is permitted. Microbiology Manager or delegate will evaluate all result in order to determine which results will be reported.

 

5.4.3      Endotoxin Samples – Refer to test specific procedure in relation to the number of re-tests to be performed.

5.4.4      Where original result has been invalidated (Phase 1), re-test of original sample, sample should be conducted.

5.4.5      Where original result is deemed valid, additional investigational testing may be permitted.

5.4.5.1    If retest of both original and retain samples confirm the original result, report the highest value obtained (i.e. worse case)

5.4.5.2    If any of the following scenarios occurs, all results should be documented on Form 690 (Phase 2, Section C); no further testing is permitted. Microbiology Manager or delegate will evaluate all result in order to determine which results will be reported.

  • Retest of original sample confirm original result but retain sample retest results are different from original result.
  • Retest results of both original and retain samples are different from original results.
  • Retest of retain sample confirm original result but original sample retest results are different from original result.

 

5.4.6      Sterility Test – Retesting (using additional samples or retain samples) is only permitted where there is unequivocal evidence which invalidates the sterility test. Refer to MICLAB 060 for further guidance.

Note: If there is not a sufficient quantity remaining from the original sample for additional testing, new samples should be selected by sampling throughout the batch using the original sample plan.

6. Appendix 1 – Flow Chart for Processing Microbiology Laboratory OOS/OOL Investigations

7. Summary of Changes

Version #Revision History
MICLAB 110New

Phase 1

Out of Specification Result

(Action Level Excursions)

o

Out of Limits Result

(Alert Limit Excursions)

o

PART A – Product or Sample Details

Form Initiated by: Unique Identifier 
Product Description DR Number 
Batch Production Number (BPN) Product Code 
Original Test Result Limits/ Specification 
Micro Manager NotifiedYes o       No oBy who?Initial and Date
     

PART B – Evaluation of Laboratory Testing

Test Type(tick)SOP Reference

Control Method reference

(if applicable)

Non-Sterile TestingoMICLAB 075 
WateroMICLAB 055 
Sterility TestingoMICLAB 060 
EndotoxinKCAoMICLAB 085 
GELoMICLAB 080 
Other ………………………..o  
Name of Technician who performed the test Training records completeYes o       No o
Date Sampled Date Test performed 
Was test conducted in accordance with SOP & Control methodYes o   No oIf No, comment:
Media/Reagents UsedLot NumberExpiry DatePassed QC Checks
   Yes o         No o
   Yes o         No o
   Yes o         No o
   Yes o         No o
   Yes o         No o
   Yes o         No o
           

PART B – Evaluation of Laboratory Testing (continued)

Equipment UsedCalibration Due DateTemperature Trends in range
  Yes o   No o             N/A o
  Yes o   No o             N/A o
  Yes o   No o             N/A o
  Yes o   No o             N/A o
  Yes o   No o             N/A o
  Yes o   No o             N/A o
Test result record Attached?Yes o   No oComments:
Are Calculations Verified and correct?Yes o   No oComments:
Negative Controls passed?Yes o   No oComments:
Other tests from same test session within limits?Yes o   No oComments:
Other Sources of Error? Technician observations during testing?Yes o   No o

Comments:

 

Brief Description of the Investigation Findings to date. 

Is initial Result Valid?

(ie has root cause been identified)

Yes o   No oIf No, is re-testing requiredYes o    No o        N/A o
If Yes, is confirmatory or investigational testing required?Yes o    No o        N/A o
If Yes, DR raised & recorded in Section AYes oInitials & Date
       

Note : If any of the above questions were answered NO, an evaluation of the validity of the test needs to be conducted by the Micro Manager.

 PrintSignDate

Approved By:

Micro Manager

   

NOTE: Refer to MICLAB 110 for appropriate Retest Procedures.

Phase 2

PART C – Retest Protocol

Retest testing of Original Sample(s)Yes o    No o             N/A o
Confirmatory Testing of Original or retention Sample(s) for investigational purposesYes o    No o             N/A o
Retest of Retain samplesYes o    No o             N/A o
Retest of Both Original and retain sampleYes o    No o             N/A o
Number of Replicates to be Tested 
Test Method 
Limits for Retest 
Retesting Conducted byPrintSign and Date
Please document the purpose of additional testing as detailed above.

Retest Protocol Approved By:

Micro Manager

PrintSignDate
   
      

PART C – Retest Protocol (continued) – Retest Results

Document result of Retesting in this section. Please attach hardcopy of raw data.

Approved By:

Micro Manager

PrintSignDate