Department | Laboratory | Document no | LAB-010 | ||
Title | Calibration Policies for Laboratory Instruments | ||||
Prepared by: | Date: | Supersedes: | |||
Checked by: | Date: | Date Issued: | |||
Approved by: | Date: | Review Date: |
Document Owner
Laboratory Manager
Affected Parties
All Laboratory/QA staffs.
Purpose
To control the Calibration for Measuring and Test Equipment, in Laboratory.
Scope
Responsibility for ensuring the procedures set out in this SOP is with the following personnel:
a. Laboratory Managers
b. Laboratory staffs
c. Engineering Services – Calibration
Definition
N/A |
Related Documents
Form-245 | Laboratory Master List of Instruments |
QMS-035 | Deviation Report System |
LAB-055 | Laboratory Results-Out Of Specification Investigation |
VAL-025 | Validation of Laboratory Instruments |
EHS Statement
Refer to safety instructions in relevant equipment manuals.
Procedure
1. Introduction
1.1. The control of calibration for measuring and testing equipment ensures that the measurement uncertainty is known and is consistent with the required measurement capability.
1.2. Laboratory equipment and instrumentation is handled and stored in such a manner so as to maintain its accuracy and fitness for use. Any special handling or storage requirements are defined in the SOP or Equipment Log applicable to the piece of equipment or instrument.
1.3. Data relating to the calibration of Laboratory inspection, measuring and testing equipment, also including records and procedures are maintained by the Laboratory.
1.4. The Engineering Services Department Calibration staff also maintains records of the Laboratory instruments that they are responsible for having calibrated (eg: balances, thermometers, oven, digital vernier, thickness indicator, incubators, air samplers, autoclave etc).
These records are kept on the Plant Maintenance Module system.
1.5. A comprehensive list of all Laboratory equipment that requires calibration is maintained by the Laboratories (Form-245).
1.6. The Laboratory master list of instruments (Form-245) and the specific requirements is kept in the Lab Equipment Log folders.
2. Control Procedures
2.1. All pieces of equipment used for testing that require calibration, are identified by a specific number or name.
2.2. The unique identifying specific number or name appears on the calibration documentation matrix relevant to the equipment or instrumentation.
2.3. In the laboratory the unique number or name is listed in the Master list of equipment and is stored in the Equipment Log.
3. Purchasing
The equipment or instrumentation is purchased using the following criteria:
3.1. Capability, (Precision, Accuracy) of the equipment.
3.2. Environmental conditions required for the equipment.
3.3. Handling and maintenance of equipment.
3.4. Calibration methods and frequency noted in SOP’s relevant to the equipment or instrumentation.
3.5. Equipment and instrumentation to be calibrated and its location.
4. Labelling
4.1. Equipment or instruments on the calibration programme are labelled when out of calibration.
4.2. In the Laboratory each piece of equipment or instrument is labelled to indicate the frequency of its calibration. The due date of calibration is noted in the logbook attached to each instrument or piece of equipment, together with the status of calibration. Equipment or instruments not requiring calibration are labelled accordingly. The calibration programme for each piece of equipment is also outlined in the Equipment Log.
4.3. When out of calibration, the QA Manager and the Laboratory Manager are notified and a label stating its status will be attached to the equipment or instrumentation.
4.4. Equipment that is out of calibration may not be used for testing. Refer to Section 6.
5. Security
5.1. A security password is used to safeguard, where applicable, software calibration settings.
5.2. The Laboratory instrument and equipment calibration is a delegated job controlled through the work rotation programme.
5.3. The Engineering Services calibration programme is controlled through the Plant Maintenance Module system (see SOP LAB-025).
5.4. The standards used for calibration are traceable to national/international standards where appropriate and are subjected to the defined storage and in use controls.
5.5. Records of all calibration activities are retained, either electronically or manually in Logbooks, by either the Engineering Services Department Calibration staffs or the Laboratory.
5.6. Refer to the Laboratory Equipment Log folders on the storage of the calibration and service records.
6. Failed Calibration
6.1. The following action is to be taken when a calibration fails on equipment or instrumentation and product has been passed in the previous period prior to the failure.
6.2. A service call is placed and the instrument or equipment is repaired and calibrated by the service technician.
6.3. The instrument is re-calibrated by a Technician according to requirements. The instrument may need to be re-validated by the manufacturer of the instrument (see SOP VAL-025).
6.4. A decision on the planned action is taken in consultation with the Analyst, QA and Laboratory Manager.
6.5. An out of specification investigation is undertaken using SOPs D.9.16. and QMS-035 where the appropriate action to investigate the effect of the un-calibrated instrument or equipment on the product is decided by the QA Manager.
6.6. If an instrument is out of calibration at the time of performing the testing, a further system is in place to assure the results by using validated methods, equipment and instruments, tight release limits, and by evaluating finished goods, stability and trials in the same testing runs an out of calibration instrument is highlighted at the point of testing.
7. Summary of Changes
Version # | Revision History |
LAB-010 | New |